Symptom Check
RAW, temperature and EC all unavailable Power, common ground, actual sensor pinout, SDI-12 address, selected GPIO, ESP-IDF and pinned half-duplex UART component. Confirm sensor variant is SDI-12, not RS485
RAW works but headline VWC unavailable Calibration status. A/B need valid weighed references and separation, C needs an independent passing check, current RAW must lie inside A/B, Calibration mode must be off and three new samples must have arrived
Wet index available but VWC unavailable Expected when only a wet reference has been saved. An index of 100 is not 100% VWC
Capture not ready Turn Calibration mode on; wait for ten fresh samples. Check RAW spread, contact, cable movement, uneven wetting and continuing drainage
Same reading for many minutes Check RAW sample age. Fresh unchanged replies are valid; a static value alone is not proof of failure
One field remains available while another is unavailable RAW, EC and temperature have independent validity/timeout checks; inspect the failing field and its protocol mapping
C check fails Verify tare, sample volume, density/packing, actual independent weight, stable moisture distribution and same probe position. Do not enter the predicted value as C
Cube reads much lower than slab Check hydraulic contact, height gradient, calibration and location. A connected upper cube cannot be assigned a fixed offset to a slab
Wet reference exceeds 100 A wetter response or changed context; the index is intentionally not clipped. Inspect sensor position, wetting, EC and saved reference
Generic VWC unavailable at some RAW values The generic curve is outside 0–100 or RAW is invalid. It is withheld instead of clamped. Calibration is not a fix for air gaps or the wrong protocol
Pore EC unavailable Experimental model starts off, requires valid checked VWC and has moisture/denominator/range gates. Use bulk EC and independent solution measurements
Trends vanish after restart or calibration Expected. Trend history does not span different calibration scales or data gaps
Wetting counts differ from irrigation controller shots They are inferred VWC-rise events. Small/overlapping shots, drainage and redistribution can merge or hide events; verify actual delivery separately
Factory node cannot be found by the generic hostname Name adds a MAC suffix in factory firmware. Use DHCP/client discovery or the actual advertised hostname

Use CONFIG.md for version and sampling settings, CALIBRATION.md for the procedure and MIGRATION-v3.md for changed entity meanings.